Choosing a thermocouple amplifier

Why the thermocouple is amplified at the sensor, and the five questions that pick the amplifier.

  • temperature
  • thermocouple
  • slip rings

The essential

A thermocouple delivers a few millivolts. An amplifier at the measurement point provides the cold junction compensation and raises the signal to a high-level voltage, linear on some models, before it crosses a slip ring or a long cable. Five questions then pick the model: mounting, number of channels, thermocouple type, range and accuracy.

01

Why amplify at the sensor?

Because the raw signal can be smaller than the noise it picks up on the way to the recorder. The thermocouple voltage is proportional to the temperature difference between the measurement junction and the cold junction, and it is in the millivolt range.

  • In a Michigan Scientific measurement at 25 °C, an unamplified type K signal of about 1 mV sat in about 30 mV of noise. Amplified, the same measurement gave about 125 mV of signal for about 25 mV of noise.
  • Cold junction compensation: every Michigan Scientific thermocouple amplifier provides it, so the output is an absolute temperature.
  • Linear output: the thermocouple voltage is not linear with temperature. An amplifier with a linear output removes the polynomial conversion from post-processing.
02

Why does it matter through a slip ring?

Because a temperature gradient across the slip ring otherwise enters the measurement. The rings and brushes are not thermocouple alloys: a temperature difference between the rotor and stator connections adds an error at these dissimilar-metal contacts. An amplifier on the rotating side holds the cold junction compensation and sends a high-level voltage through the rings, which the gradient does not affect.

03

How is it mounted?

On the part itself. The amplifiers are adhered, potted or strapped to the part, including a rotating one.

  • MicroTC: 32.0 × 17.4 × 8.2 mm, 3.9 g. AMP-TC-M1: 30 × 20 × 6 mm.
  • AMP-TCN-X2: a Ø 53.37 × 21.34 mm module that pilots on the rotor of SR series slip rings; modules stack for more channels.
  • Short S-Series slip rings exist with the thermocouple amplifier built into the rotor (S6/TC3, S8/TC4).
04

How many channels, linear or not?

The amplifiers with a linear output have one channel. For several channels in one unit, the output is not linearized and is converted with the polynomials given in the manual.

  • Linear, one channel: MicroTC, 5 mV/°C, 2.35 kHz bandwidth. AMP-TC-LIN, 10 mV/°C, 10 Hz bandwidth.
  • Not linearized, 10 mV/°C at 25 °C: AMP-TC-M1 with 1, 2 or 3 channels, 1.16 kHz; AMP-TCN-X2 with 1 to 9 channels, 1.56 kHz.
  • Several amplifiers can run from one Michigan Scientific control unit.
05

Which thermocouple type and range?

The thermocouple type narrows the choice first, then the range to be read without polynomial correction.

  • MicroTC: type K only. Linear from −25 to 400 °C; −200 to 970 °C with the post-processing polynomials.
  • AMP-TC-LIN: types K and J. Linear from −80 to 1 000 °C with type K, 0 to 750 °C with type J.
  • AMP-TC-M1: types K, J, T and E. AMP-TCN-X2: types K, J and T, others on request. Range by thermocouple type, through the polynomials of the manual, for example 0 to 1 370 °C with type K.
06

What accuracy can be expected?

± 2 °C to ± 3 °C near room temperature, cold junction compensation included, and ± 5 °C at the ends of the case temperature range of the amplifier. Each datasheet gives the limits.

  • MicroTC: ± 2 °C typical, ± 3 °C max.
  • AMP-TC-M1: ± 2 °C at 25 °C case temperature, ± 3 °C from −25 to 85 °C, ± 5 °C from −55 to 125 °C.
  • AMP-TCN-X2: ± 1 °C at 25 °C, ± 2 °C from −25 to 85 °C, ± 5 °C from −55 to 125 °C.
  • AMP-TC-LIN: ± 3 °C from −40 to 70 °C case temperature, ± 5 °C up to 110 °C.

To remember

  • Amplify at the sensor, above all through a slip ring.
  • One linear channel: MicroTC (type K) or AMP-TC-LIN (K, J).
  • Several channels, or types T and E: AMP-TC-M1 (1 to 3) or AMP-TCN-X2 (1 to 9), with the polynomials.
  • ± 2 to ± 3 °C near room temperature, ± 5 °C at the ends of the case temperature range.

A temperature to measure on a rotating part?

Describe the part, the range and the number of points. An engineer answers with the amplifier and the way the signal leaves the part.

More details to give? Use the full contact form.